Technical Date

L8P  fetal/maternal monitor

Fetal Heart Rate:

Transducer: Multi-crystals, pulsed doppler , high sensitively 

Strength: <5mW/cm2

Working frequency: 1.0MHz

Signal processing:    special DSP system and modern recognization.

Measurement range:  50~210 bpm

Alarm Range:

High limit:  160,170,180,190 bpm

Low limit: 90,100,110,120 bpm

Maximum audio output power : 1.5 Watt

Toco:

Measurement range:  0~100 units

Maternal SpO2 measure:

Measure scope : 70%~99%.

Measure accuracy: ±3%

Pregnant HR Measure:

Measure scope: 30bpm~240bpm

Measure accuracy: ±2 bpm

NIBP measure:

a)   NIBP-SYS (6.7~32.0Kpa)(50~240mmHg)

                  b)  NIBP-DIA (3.4~26.6Kpa)(25~200mmHg)

c)       NIBP-MEAN(2.0~24.0ka(15~180mmHg)

NIBP Accuracy: ±1.1kPa(±8mmHg)or ±5% of the results, take the bigger one.

Maximum Mean error : ±5mmHg

    Maximum Standard deviation 8mmHg

Pulse Rate Accuracy: Maximum Mean error : ±2bpm 

Measure accuracy: ±2 bpm or ±5% of the results, take the bigger one.

Measure mode: Manual

ECG

CMRR: ≥90dB

Heart Rate Measurement and Alarm Range: 15-300bpm

Accuracy: ±1bpm

Heart Rate Average: 8 beats

Lead Selection: II

Input: 5-lead ECG cable

Sweep Speed: 25 mm/s

Amplitude Selection: x1/4, x1/2, x1, x2, x4

Frequency Response:

Monitoring:   0.5 to 35Hz (+0.4dB, 3.0dB)

Defibrillation Protection: 5kV test

Recovery Time Following Defibrillation: Less than 5 seconds

Temperature:

Measurement and Alarm Limit: 0-50°C

Probe: Skin surface

Unit: Celsius/Fahrenheit

Accuracy: ±0.1°C(25°C ~45°C), ±0.2°C(0°C ~25°C or 45°C ~50°C)

Resolution: 0.1°C

Respiration

Measurement Mode: Thoracic Impedance

Respiration Rate: 0-150 bpm

Waveform Display Speed: 25mm/s

 

Display:

The LCD displays FHR, TOCO, FM , maternal parameters , time, date, volume and so on, It can store and play

back the data.

Power: a.c.100-240v, 50/60Hz or 16.8v Li-ION battery

Power Consumption: < 20W

Dimension: 350W×320D×85H(mm)

Net weight:3.5kg

Packing size:41.3×39-×24.5 (mm)

Gross Weight:9Kg

Environment

    Working environment: Temperature: +5°C ~ +40°C

Humidity: < 85%

                         Atmospheric pressure: 86kPa ~ 106kPa

Transport and storage temperature: Temperature: -10°C ~ +55°C

                            Humidity: < 93%

                         Atmospheric pressure: 86kPa ~ 106kPa

Transducer Acoustic Output:

Under the requirements laid down in IEC 1157, 1992, the peak negative acoustic pressure

does not exceed 1Mpa. The output beam intensity does not exceed 20mW/cm2

and the spatial-peak temporal-average intensity does not exceed 100mW/cm2

GEL:

Viscous aqueous non-sensitizing , hypo-allergenic and non-irritating to skin. Indefinite shelf life,

Bacteriostat (not sterile).

                   

3.1 Unpacking the units

3.1.1 Carefully remove the main unit and accessories from the carton.

3.1.2 Check that the content of the carton corresponds with the list below and that all items are undamaged .

         1 main unit of fetal monitor

                                              3 in 1 transducer(ultrasound transducer ,contraction transducer, remote event marker)

                                             2 transducer belts

         1 mains power cord and adaptor

         2 paper pads

         Warranty card

         1 250ml bottle of gel

         Use manua

        One Spo2 sensor(option)

     One blood pressure cuff(option)

 

 

 

 

 

 

 